Temperature equipment |
Sample stage sensor/heater power: | DT-670-CU-HT calibrated silicon diode/100 W |
Radiation shield sensor/heater power: | DT-670C-CU silicon diode/100 W |
Probe arm: | DT-670C-SD silicon diode (on one probe arm)/no active control |
Electronics console: |
Temperature controllers: | Two Model 336s |
Helium transfer line: | Large bore S-style helium/nitrogen transfer line with flow control 12.7 mm (0.5 in) top port withdrawal |
| |
Included sample holder: | Removable SH-4.00-G, 102 mm (4 in) maximum sample size, grounded |
| |
Station footprint: | 1067 mm (42 in) width × 851 mm (33.5 in) depth (does not include console) |
| |
Vacuum chamber |
Material: | Electroless nickel-plated aluminum |
Overall size: | 406 mm (16 in) outer diameter × 178 mm (7 in) tall |
Chamber access: | 330 mm (13 in) diameter opening |
Viewport: | 102 mm (4.0 in) diameter fused quartz |
Viewing area: | 91 mm (3.6 in) diameter |
Pump port: | NW 40 with included vacuum isolation valve |
Gauge port: | NW 25 |
Gas port: | 0.25 in NPT with valve |
Over-pressure safety: | 3.5 kPa (0.5 psi) pressure relief valve |
| |
Radiation shielding |
Material: | Electroless nickel-plated aluminum |
Sample access: | 180 mm (7.1 in) diameter lid |
Viewport: | 159 mm (6.25 in) diameter IR absorbing window |
Viewing area: | 152 mm (6.0 in) diameter |
| |
|
| |
Included kits |
Tool kit: | Included tools for standard operation |
Spares kit: | Included kit for common consumable items |
Probe starter kit: | ZN50R-25-W (qty 2), ZN50R-25-BECU (qty 2), and ZN50R-CVT-25-W (qty 2) for training |
| |
Utilities |
1-phase voltage: | 100/120/220/240 VAC (+5%, -10%), 50/60 Hz |
1-phase power: | 2 kVA recommended |
3-phase voltage: | Not required |
3-phase power: | Not required |
Cooling water: | Not required |
| |
Approval: | All instruments CE marked |