Output selection modes | Voltage or current |
Range selection modes | Manual, autorange |
Current source compliance selection | 10 V or 100 V |
Programmable limits (DC only) | I out: compliance voltage, V out: current limit |
Accuracy | Positive output: +10%/-0% of setting (when setting is ≥10% of its full-scale range); Negative output: -10%/+0% of setting (when setting is ≥10% of its full-scale range) |
Programming resolution | Current limit: 0.1% of full-scale current range (auto selected) Voltage limit: 10 mV (10 V compliance), 100 mV (100 V compliance) |
Polarity | Bipolar, 4-quadrant |
Load impedance | Capable of driving any resistive load within the I and V limitations of the source; stability maintained with reactive loads up to 50 µF or 1 mH (with 100 Ω damping) |
AC frequency range | 100 mHz to 100 kHz |
AC frequency accuracy | ±0.06% |
AC signal types | Sine |
Total harmonic distortion | <0.1% from DC to 20 kHz, <1% from 20 kHz to 100 kHz |
Sample rate | 600 kilosamples per second |
Noise (10 Hz to 40 MHz)2,5 | Voltage: ≤700 µV RMS, across all ranges Current: ≤8 µA RMS (400 µV RMS across 50 Ω), across all ranges |
Load regulation2 | Voltage: 10 V range and below: <0.2% of range, 100 V range: <0.01% of range; Current: 10 V compliance: <0.01% of range, 100 V compliance: <0.01% of range |
Settling times | Voltage: time to reach final value at open load condition: 25 µs
Current: time to reach final value into a given load4: 25 µs
|
Output lead connections | Output HI, Output LO, Guard, Chassis Ground |
Connector type | Front: safety banana jack Rear: BNC or triaxial (user selectable between front and rear panel) |
Guard output | Maximum load current: 1 mA peak |
Power up protection | Output terminals floating on power up
|
Additional output features | User-specified output setting limit, enable/disable output setting
|
Warm-up time | 30 min (to achieve specified accuracy) |
Isolation | Source output isolated from chassis ground
|
Output capacitance | Front: <100 pF, rear: <50 pF |