lipix

Transverse 400 Series Hall Probe Specifications

Click here for Hall probe active areas

By default, each probe has a 2 m long cable (except for Model 410 and cryogenic probes). Also available: 6 m, 10 m, and 30 m lengths.

Transverse probes

Transverse probe

Operating temperature range: 0 °C to +75 °C

 RoHS compliant
L mm (in)T mm (in)W mm (in)A mm (in)Active area mm (in)Stem materialFrequency rangeUsable full scale rangesCorrected accuracy (% rdg at 25 °C) Temp coefficient (max) zeroTemp coefficient (max) of sensitivityContains temp sensor
for Models 475, 455, and 425
HMMT-6J02-VRNo50.8 ±3.2 (2 ±0.125)1.55 (0.061) max4.57 ±0.20
(0.180 ±0.008)
3.81 ±1.27
(0.150 ±0.050)
Approx 1.02 dia (0.040 dia)AluminumDC to 800 HzHSE
3.5 G, 35 G, 350 G, 3.5 kG, 35 kG
±0.20% to 30 kG and ±0.25% 30 to 35 kG±0.09 G/°C455/475: ±0.015%/°C

425: -0.04%/°C
Yes
HMMT-6J04-VRYes101.6 ±3.2 (4 ±0.125)
HMMT-6J08-VRNo203.2 ±3.2 (8 ±0.125)
HMMT-6J12-VRNo305 ±6.4 (12 ±0.25)
HMMT-6J18-VRNo457 ±6.4 (18 ±0.25)
HMNT-4E02-VRNo50.8 ±3.2 (2 ±0.125)1.14 (0.045) max3.81 ±0.13
(0.150 ±0.005)
Fiberglass epoxyDC to 20 kHz
HMNT-4E04-VRYes101.6 ±3.2 (4 ±0.125)
HMMT-6J02-VFNo50.8 ±3.2 (2 ±0.125)1.55
(0.061) max
4.57 ±0.20
(0.180 ±0.008)
AluminumDC to 400 HzHST-4
35 G, 350 G, 3.5 kG, 35 kG
±0.10% to 30 kG and ±0.15% 30 to 35 kG±0.13 G/°C-0.005%/°C
HMMT-6J04-VFYes101.6 ±3.2 (4 ±0.125)
HMMT-6J08-VFNo203.2 ±3.2 (8 ±0.125)
HMMT-6J12-VFNo305 ±6.4 (12 ±0.25)
HMMT-6J18-VFNo457 ±6.4 (18 ±0.25)
HMNT-4E02-VFNo50.8 ±3.2 (2 ±0.125)1.14 (0.045) max3.81 ±0.13
(0.150 ±0.005)
Fiberglass epoxyDC to 800 Hz
HMNT-4E04-VFYes101.6 ±3.2 (4 ±0.125)
 RoHS compliant
L mm (in)T mm (in)W mm (in)A mm (in)Active area mm (in)Stem materialFrequency rangeUsable full scale rangesCorrected accuracy (% rdg at 25 °C)Temp coefficient (max) zeroTemp coefficient (max) of sensitivityContains temp sensor
for Models 460, 450, and 421
MMT-6J02-VHNo50.8 ±3.2 (2 ±0.125)1.55
(0.061) max
4.57 ±0.20
(0.180 ±0.008)
3.81 ±1.27
(0.150 ±0.050)
Approx 1.02 dia (0.040 dia)AluminumDC and 10 Hz to 100 HzHSE-1
30 G, 300 G, 3 kG, 30 kG
±0.25% to 30 kG±0.09 G/°C450/460: ±0.015%/°C

421: -0.04%/°C
Yes
MMT-6J04-VH101.6 ±3.2 (4 ±0.125)
MMT-6J08-VH203.2 ±3.2 (8 ±0.125)
MMT-6J12-VH305 ±6.4 (12 ±0.25)
MMT-6J18-VH457 ±6.4 (18 ±0.25)
MNT-4E02-VH50.8 ±3.2 (2 ±0.125)1.14 (0.045) max3.81 ±0.13
(0.150 ±0.005)
Fiberglass epoxyDC and 10 Hz to 400 Hz
MNT-4E04-VH101.6 ±3.2 (4 ±0.125)
MMT-6J02-VG50.8 ±3.2 (2 ±0.125)1.55 (0.061) max4.57 ±0.20
(0.180 ±0.008)
AluminumDCHST-2
300 G, 3 kG, 30 kG
±0.15% to 30 kG±0.13 G/°C-0.005%/°CNo
MMT-6J04-VG101.6 ±3.2 (4 ±0.125)
MMT-6J08-VG203.2 ±3.2 (8 ±0.125)
MMT-6J12-VG305 ±6.4 (12 ±0.25)
MMT-6J18-VG457 ±6.4 (18 ±0.25)
MNT-4E02-VG50.8 ±3.2 (2 ±0.125)1.14 (0.045) max3.81 ±0.13
(0.150 ±0.005)
Fiberglass epoxyDC to 10 Hz to 400 Hz
MNT-4E04-VG101.6±3.2 (4±0.125)

Transverse probes for cryogenic applications

Cryogenic transverse probes


 RoHS compliant
L mm (in)D mm (in)A mm (in)Active area mm (in)Stem materialFrequency rangeUsable full scale rangesCorrected accuracy (% rdg at 25 °C) Operating temp range Temp coefficient (max) zeroTemp error (approx) calibrationContains temp sensor
for Models 475, 455, and 425300 Kref 
HMCT-3160-WNNo1549 ±25.4 (61 ±1)6.35 dia ±0.25 (0.25 dia ±0.010)5.33 ±1.27 (0.210 ±0.050)Approx 1 dia (0.040 dia)Stainless steelDC to 800 HzHST-3
35 G, 350 G, 3.5 kG, 35 kG, 350 kG
±2% to 100 kG1.5 K to 350 K±0.13 G/°C200 K+0.05%No
100 K-0.04%
80 K-0.09%
for Models 460, 450, and 42120 K-0.40% 
MCT-3160-WNNo1549 ±25.4 (61 ±1)6.35 dia ±0.25 (0.25 dia ±0.010)5.33 ±1.27 (0.210 ±0.050)Approx 1 dia (0.040 dia)Stainless steelDC and 10 Hz to 400 HzHST-1
300 G, 3 kG, 30 kG, 300 kG
±2% to 100 kG1.5 K to 350 K±0.13 G/°C4 K-0.70%No
1.5 K-1.05%