Tangential 400 Series Hall Probe Specifications
Click here for Hall probe active areas
By default, each probe has a 2 m long cable (except for Model 410 and cryogenic probes). Also available: 6 m, 10 m, and 30 m lengths.
Tangential probes
Operating temperature range: 0 °C to +75 °C
L mm (in) | T mm (in) | W mm (in) | A mm (in) | Active area mm (in) | Stem material | Frequency range | Usable full scale ranges | Corrected accuracy (% rdg at 25 °C) | Temp coefficient (max) zero | Temp coefficient (max) of sensitivity | Contains temp sensor | ||
for Models 475, 455, and 425 | |||||||||||||
HMNTAN-DQ02-TH | No | 38.1 ±1.52 (1.5 ±0.060) | 2.8 (0.11) max | 6.6 (0.26) max | 0.76 ±0.18 (0.030 ±0.005) | Approx 0.51 dia (0.020 dia) | Plastic | DC to 400 Hz | HSE 3.5 G, 35 G, 350 G, 3.5 kG, 35 kG | 0.25% to 20 kG | ±0.1 G/°C | 455/475: ±0.02%/°C 425: -0.05%/°C | Yes |
L mm (in) | T mm (in) | W mm (in) | A mm (in) | Active area mm (in) | Stem material | Frequency range | Usable full scale ranges | Corrected accuracy (% rdg at 25 °C) | Temp coefficient (max) zero | Temp coefficient (max) of sensitivity | Contains temp sensor | ||
for Models 460, 450, and 421 | |||||||||||||
MNTAN-DQ02-TH | No | 38.1 ±1.52 (1.5 ±0.060) | 2.8 (0.11) max | 6.6 (0.26) max | 0.76 ±0.18 (0.030 ±0.005) | Approx 0.51 dia (0.020 dia) | Plastic | DC and 10 Hz to 400 Hz | HSE-1 30 G, 300 G, 3 kG, 30 kG | 0.25% to 20 kG | ±0.1 G/°C | -0.05%/°C | No |