lipix

Multi-axis 400 Series Hall Probe Specifications

Click here for Hall probe active areas

By default, each probe has a 2 m long cable (except for Model 410 and cryogenic probes). Also available: 6 m and 10 m lengths.

Three-axis probes

Three-Axis Probes

 L mm (in)Stem materialFrequency rangeUsable full-scale rangesCorrected accuracy (% rdg at 25 °C)Operating temp rangeTemperature coefficient (max) zeroTemperature coefficient (max) of sensitivityContains temperature sensor
for Model 460
MMZ-2502-UH54±3.2 (2.125±0.125)AluminumDC and 10 Hz to 400 HzHSE-1
30 G, 300 G, 3 kG, 30 kG
0.25% to 20 kG; 0.5% from 20 kG to 30 kG10 °C to 40 °C±0.09 G/°C±0.015%/°CYes
MMZ-2504-UH104.8±3.2 (4.125±0.125)
MMZ-2508-UH206.4±3.2 (8.125±0.125)
MMZ-2512-UH308±3.2 (12.125±0.125)
MMZ-2518-UH460±6.4 (18.125±0.25)
MMZ-2536-UH918±6.4 (36.125±0.25)
MMZ-2560-UH1534±6.4 (60.375±0.5)
The sensors at the ends of the multi-axis probes are quite fragile and susceptible to damage. Lake Shore offers an assortment of brass covers for probe protection during use with DC fields. Consult Lake Shore for model numbers and pricing.


Two-axis probes (DISCONTINUED)

Two-Axis Probes

 L mm (in)Stem materialFrequency rangeUsable full-scale rangesCorrected accuracy (% rdg at 25 °C)Operating temp rangeTemperature coefficient (max) zeroTemperature coefficient (max) of sensitivityContains temperature sensor
for Model 460
MMY-1802-UH50.8 ±3.2
(2 ±0.125)
AluminumDC and 10 Hz to 400 HzHSE-1
30 G, 300 G, 3 kG, 30 kG
0.25% to 20 kG; 0.5% from 20 kG to 30 kG10 °C to 40 °C±0.09 G/°C±0.015%/°CYes
MMY-1808-UH203.2 ±3.2
(8 ±0.125)
MMY-1818-UH457 ±6.4
(18 ±0.25)
MMY-1836-UH914 ±6.4
(36 ±0.25)