Temperature options | Standard room temperature Low temp 10 K to 350 K (optical CCR), 400 K (CCR) High temp room temperature to 1273 K (oven) |
---|
Sample type | van der Pauw, Hall bar |
---|
Maximum sample size | 10 mm × 10 mm × 3 mm |
---|
Software | MeasureLINK™ software included to execute Hall measurements |
---|
Gate bias option | Up to 100 V |
---|
Legacy conversion | Full legacy 8400 replacement | Partial legacy 8400 replacement |
---|
Resistance range options | Low-resistance Coming soon Standard resistance 0.5 mΩ to 10 MΩ High-resistance 10 MΩ to 200 GΩ | Standard resistance 10 mΩ to 10 MΩ High-resistance 10 MΩ to 200 GΩ |
---|
Measurement type | DC field Hall measurement AC field Hall measurement | DC field Hall measurement FastHall measurement (AC electrical, DC field) |
---|
Measurement-specific instrumentation | Base 776B and M81-SSM-6 + BCS-10 Standard resistance BCS‑10, VM‑10 High-resistance VM‑10, VS‑10, CM‑10 | Standard resistance M91 High-resistance M91-HR |
---|
DC field options (nominal) | 4-inch magnet 1.69 T room temp and 0.91 T variable temp 7-inch magnet 2.28 T room temp and 1.54 T variable temp |
---|
AC field range options (fixed-field, nominal) | Frequency: 50 mHz or 100 mHz 4-inch magnet 1.18 T (RMS) room temp and 0.63 T (RMS) variable temp 7-inch magnet 1.20 T (RMS) room temp and 0.69 T (RMS) variable temp | None |
---|
Resistivity | 10-5 to 105 Ω·cm | 10-3 to 105 Ω·cm |
---|
Carrier concentration density | 8 × 102 to 8 × 1023 cm-3 | 8 × 102 to 8 × 1023 cm-3 |
---|
Mobility range | 1 × 106 (DC field) and 10-3 to 106 (AC field) cm2/V s | 10-3 to 106 cm2/V s |
---|
Multi-carrier analysis | QMSA-compatible (DC field only) | QMSA-compatible |
---|