ULTRA-LOW VIBRATION
Add on the low-vibration option to reach ultra-low vibrations with less than 2 nm vibration (>5 Hz).
STABLE TEMPERATURE
Sensitive samples require a stable environment. Guard against thermal-induced changes with low thermal drift. Precise to 15 mK at base temperature.
AUTOMATED SAFETY ALERTS
- Water shut off
- Loss of power
- Contaminated helium
- Small leak
- Major leak
- Equipment not responding
- Helium pressure out of range
- Maintenance due
VERSATILE
Easily move Infinite Helium between multiple continuous flow cryostats.
6 MONTHS+ CONTINUOUS RUN TIME
All components are selected for performance longevity. Don't interrupt your research because of servicing
FACILITY REQUIREMENTS
Recommended compressor maintenance interval 30,000 h
Recommended cold head maintenance interval 10,000 h
Water-cooled
60 Hz power requirements | 50 Hz power requirements | Cooling water requirements | Compressor size |
---|
200 VAC, 3-phase, 7.5 to 7.8 kW or 480 VAC, 3-phase, 7.5 to 7.8 kW | 200 VAC, 3-phase, 6.6 to 6.9 kW or 380 to 415 VAC, 3-phase, 6.6 to 6.9 kW | 6 to 9 L/min at 5 to 25 °C | 443 mm × 493 mm × 532 mm high; 100 kg |
Air-cooled
60 Hz power requirements | 50 Hz power requirements | Cooling air requirements | Compressor size |
---|
200 VAC, 3-phase, 7.5 to 8.3 kW steady state or 460/480 VAC, 3-phase, 7.5 to 8.3 kW | 200 VAC, 3-phase, 6.5 to 7.2 kW steady state or 380/400/415 VAC, 3-phase, 6.5 to 7.2 kW | 23 m3/min | 450 mm × 485 mm × 925 mm high; 155 kg |