SuperTran-VP cryostat specifications

STVP-100 Series specifications

 STVP-100STVP-100-TH
Cooling typeLHe
Temperature range<2 K to 325 K (LHe); 65 K to 325 K (LN2)<2 K to 420 K (LHe); 65 K to 420 K (LN2)
Typical temperature stability1±50 mK
Sample environmentFlowing vapor
Sample typeAll
Cooldown30 min (to 5 K)
Optical
Rotatable sample
Orientation2Vertical for operation <4.5 K
Cryogen consumption (LHe room temp to 4.2 K)0.5 L
Cryogen consumption (LHe at 5 K)1.3 L/h
Height (approximate)762 mm (30 in)
Inner space (at sample region)38.1 mm (1.5 in)
Sample mount diameter31 mm (1.25 in)
Window block82.6 mm (3.25 in) square
Weight (excluding transfer line, approximate)7 kg (15.4 lb)
Shipping weight (cryostat + line, approximate)61 kg (135 lb)
Shipping dimensions (cryostat + line, approximate)1905 × 990.6 × 431.8 mm (75 × 39 × 17 in)
1Measured with temperature controller
2Cryogen consumption may be higher during non-vertical operation

 

STVP-NMR specifications

 STVP-NMR
Cooling typeLHe
Temperature range<2 K (1.5 K in single-shot mode) to 325 K (420 K optional)
Typical temperature stability1±50 mK
Sample environmentFlowing vapor
Sample typeAll
Cooldown30 min (to 5 K)
Orientation2Vertical for operation <4.5 K
Cryogen consumption (LHe room temp to 4.2 K)0.5 L
Cryogen consumption (LHe at 5 K)1.4 L/h
Height (approximate)Customer-specified to fit magnet
Inner space (at sample region)55 mm (2.16 in)
Weight (excluding transfer line, approximate)7 kg (15.4 lb)
Shipping weight (cryostat + line, approximate)79 kg (174 lb)
Shipping dimensions (cryostat + line, approximate)1905 × 990.6 × 431.8 mm (75 × 39 × 17 in)
1Measured with temperature controller
2Cryogen consumption may be higher during non-vertical operation

 

Please see the FTIR datasheet for STVP-FTIR specifications