lipix

CCR vacuum cryostat specifications

 CCS-100CCS-300SCCS-300STCCS-XGCCS-400CCS-400H
Cooling typeCCR
Temperature range<4 K to 325 K<4 K to 325 K<4 K to 325 K<4 K to 325 K<4 K to 500 K<5 K to 800 K
Typical temperature stability1±50 mK
Sample environmentVacuum
Sample typeConductive, solid
Cooldown1 h to 2 h1.5 h to 2.5 h1.5 h to 2.5 h2 h to 3 h1.5 h to 3 h2 h to 2.5 h
Vibration<40 nm
SizeCompactCompact
Cold head locationAnyAnyAnyTopAnyAny
Optical
Height (approximate)56 to 84 cm
(22 to 33 in)
71 to 99 cm
(28 to 39 in)
71 to 99 cm
(28 to 39 in)
66 to 94 cm
(26 to 37 in)
61 to 89 cm
(24 to 35 in)
66 to 94 cm
(26 to 37 in)
Weight (approximate)16 to 29 kg
(36 to 64 lb)
17 to 30 kg
(37 to 66 lb)
17 to 30 kg
(37 to 66 lb)
17 to 30 kg
(37 to 66 lb)
16 to 30 kg
(36 to 65 lb)
16 to 30 kg
(36 to 65 lb)
Window block size83 mm to 85 mm (3.25 in to 3.75 in) square38 mm (1.5 in) square83 mm to 85 mm (3.25 in to 3.75 in) square83 mm to 85 mm (3.25 in to 3.75 in) square83 mm to 85 mm (3.25 in to 3.75 in) square
Recommended maintenance13,000 h
1Measured with temperature controller

Facility requirements

Water-cooled

CCS-Cold headRecommended compressor maintenance intervalRecommended cold head maintenance interval60 Hz power requirements50 Hz power requirementsCooling water requirementsCompressor size
100
300S
300ST
400
400H
XG
-202
-204
-204N
30,000 h13,000 h208 to 230 VAC, 1-phase, 2.6 kW200, 220 to 240 VAC, 1-phase, 2.25 to 2.4 kW1.9 to 3.8 L/min at 4 to 27 °C444 mm × 453 mm × 617 mm high; 73 kg
100
300S
300ST
400
XG
-10110,000 h208 to 230 VAC, 1-phase, 2.6 kW200, 220 to 240 VAC, 1-phase, 2.25 to 2.4 kW1.9 to 3.8 L/min at 4 to 27 °C444 mm × 453 mm × 617 mm high; 73 kg
100
300S
300ST
400
400H
XG
-408
-415
-418
200 VAC, 3-phase, 7.5 to 7.8 kW or 480 VAC, 3-phase, 7.5 to 7.8 kW200 VAC, 3-phase, 6.6 to 6.9 kW or 380 to 415 VAC, 3-phase, 6.6 to 6.9 kW6 to 9 L/min at 5 to 25 °C443 mm × 493 mm × 532 mm high; 100 kg

Air-cooled

CCS-Cold headRecommended compressor maintenance intervalRecommended cold head maintenance interval60 Hz power requirements50 Hz power requirementsCooling air requirementsCompressor size
100
300S
300ST
400
400H
XG
-202
-204
-204N
30,000 h13,000 h208 to 230 VAC, 1-phase, 2.6 kW200, 220 to 240 VAC, 1-phase, 2.25 to 2.4 kW17.6 m3/min (60 Hz) or 14.7 m3/min (50 Hz) 444 mm × 453 mm × 876 mm high; 103 kg
100
300S
300ST
400
XG
-10110,000 h208 to 230 VAC, 1-phase, 2.6 kW200, 220 to 240 VAC, 1-phase, 2.25 to 2.4 kW17.6 m3/min (60 Hz) or 14.7 m3/min (50 Hz) 444 mm × 453 mm × 876 mm high; 103 kg
100
300S
300ST
400
400H
XG
-408
-415
-418
200 VAC, 3-phase, 7.5 to 8.3 kW steady state or 460/480 VAC, 3-phase, 7.5 to 8.3 kW200 VAC, 3-phase, 6.5 to 7.2 kW steady state or 380/400/415 VAC, 3-phase, 6.5 to 7.2 kW23 m3/min 450 mm × 485 mm × 925 mm high; 155 kg