| CCS-100 | CCS-300S | CCS-300ST | CCS-XG | CCS-400 | CCS-400H |
---|
Cooling type | CCR |
---|
Temperature range | <4 K to 325 K | <4 K to 325 K | <4 K to 325 K | <4 K to 325 K | <4 K to 500 K | <5 K to 800 K |
---|
Typical temperature stability1 | ±50 mK |
---|
Sample environment | Vacuum |
---|
Sample type | Conductive, solid |
---|
Cooldown | 1 h to 2 h | 1.5 h to 2.5 h | 1.5 h to 2.5 h | 2 h to 3 h | 1.5 h to 3 h | 2 h to 2.5 h |
---|
Vibration | — | — | — | <40 nm | — | — |
---|
Size | — | Compact | Compact | — | — | — |
---|
Cold head location | Any | Any | Any | Top | Any | Any |
---|
Optical | ✔ | ✔ | ✖ | ✔ | ✔ | ✔ |
---|
Height (approximate) | 56 to 84 cm (22 to 33 in) | 71 to 99 cm (28 to 39 in) | 71 to 99 cm (28 to 39 in) | 66 to 94 cm (26 to 37 in) | 61 to 89 cm (24 to 35 in) | 66 to 94 cm (26 to 37 in) |
---|
Weight (approximate) | 16 to 29 kg (36 to 64 lb) | 17 to 30 kg (37 to 66 lb) | 17 to 30 kg (37 to 66 lb) | 17 to 30 kg (37 to 66 lb) | 16 to 30 kg (36 to 65 lb) | 16 to 30 kg (36 to 65 lb) |
---|
Window block size | 83 mm to 85 mm (3.25 in to 3.75 in) square | 38 mm (1.5 in) square | — | 83 mm to 85 mm (3.25 in to 3.75 in) square | 83 mm to 85 mm (3.25 in to 3.75 in) square | 83 mm to 85 mm (3.25 in to 3.75 in) square |
---|
Recommended maintenance | 13,000 h |
---|
1Measured with temperature controller
Facility requirements
Water-cooled
CCS- | Cold head | Recommended compressor maintenance interval | Recommended cold head maintenance interval | 60 Hz power requirements | 50 Hz power requirements | Cooling water requirements | Compressor size |
---|
100 300S 300ST 400 400H XG | -202 -204 -204N | 30,000 h | 13,000 h | 208 to 230 VAC, 1-phase, 2.6 kW | 200, 220 to 240 VAC, 1-phase, 2.25 to 2.4 kW | 1.9 to 3.8 L/min at 4 to 27 °C | 444 mm × 453 mm × 617 mm high; 73 kg |
---|
100 300S 300ST 400 XG | -101 | 10,000 h | 208 to 230 VAC, 1-phase, 2.6 kW | 200, 220 to 240 VAC, 1-phase, 2.25 to 2.4 kW | 1.9 to 3.8 L/min at 4 to 27 °C | 444 mm × 453 mm × 617 mm high; 73 kg |
---|
100 300S 300ST 400 400H XG | -408 -415 -418 | 200 VAC, 3-phase, 7.5 to 7.8 kW or 480 VAC, 3-phase, 7.5 to 7.8 kW | 200 VAC, 3-phase, 6.6 to 6.9 kW or 380 to 415 VAC, 3-phase, 6.6 to 6.9 kW | 6 to 9 L/min at 5 to 25 °C | 443 mm × 493 mm × 532 mm high; 100 kg |
---|
Air-cooled
CCS- | Cold head | Recommended compressor maintenance interval | Recommended cold head maintenance interval | 60 Hz power requirements | 50 Hz power requirements | Cooling air requirements | Compressor size |
---|
100 300S 300ST 400 400H XG | -202 -204 -204N | 30,000 h | 13,000 h | 208 to 230 VAC, 1-phase, 2.6 kW | 200, 220 to 240 VAC, 1-phase, 2.25 to 2.4 kW | 17.6 m3/min (60 Hz) or 14.7 m3/min (50 Hz) | 444 mm × 453 mm × 876 mm high; 103 kg |
---|
100 300S 300ST 400 XG | -101 | 10,000 h | 208 to 230 VAC, 1-phase, 2.6 kW | 200, 220 to 240 VAC, 1-phase, 2.25 to 2.4 kW | 17.6 m3/min (60 Hz) or 14.7 m3/min (50 Hz) | 444 mm × 453 mm × 876 mm high; 103 kg |
---|
100 300S 300ST 400 400H XG | -408 -415 -418 | 200 VAC, 3-phase, 7.5 to 8.3 kW steady state or 460/480 VAC, 3-phase, 7.5 to 8.3 kW | 200 VAC, 3-phase, 6.5 to 7.2 kW steady state or 380/400/415 VAC, 3-phase, 6.5 to 7.2 kW | 23 m3/min | 450 mm × 485 mm × 925 mm high; 155 kg |
---|