| | 231 | 231P | 234 |
Output |
| Number of outputs | 1 | 1 | 1 |
| Output type | Current source, isolated from power source—output or sensor can be grounded, but not both (all models) |
| Output range | 4 mA to 20 mA or 0 mA to 20 mA (for 0 V to 10 V with provided 500 Ω 0.02%, 25 ppm resistor) (all models) |
| Output compliance | 10 V (500 Ω max load) | 10 V (500 Ω max load) | 10 V (500 Ω max load) |
| Output temperature ranges | | Range 1 | | Range 2 | | Range 3 | | Range 4 | | Range 5 | | Range 6 |
| | | | 0 K to 20 K | | 0 K to 100 K | | 0 K to 200 K | | 0 K to 325 K | | 0 K to 475 K | | 0 K to 1000 K |
| | | | 0 K to 20 K | | 0 K to 100 K | | 0 K to 200 K | | 0 K to 325 K | | 0 K to 475 K | | 0 K to 1000 K |
| | | | 0 K to 10 K | | 0 K to 20 K | | 0 K to 100 K | | 0 K to 200 K | | 0 K to 300 K | | 75 K to 325 K |
|
| 4 mA to 20 mA output | | Output resolution | | Current | | Temperature equivalence | | Range 1 | | Range 2 | | Range 3 | | Range 4 | | Range 5 | | Range 6 |
| | | | | | 1.22 µA (0.006% of full scale) | | | | 1.5 mK | | 7.6 mK | | 15.3 mK | | 24.8 mK | | 36.2 mK | | 76.3 mK |
| | | | | | 1.22 µA (0.006% of full scale) | | | | Not used | | 7.6 mK | | 15.3 mK | | 24.8 mK | | 36.2 mK | | 76.3 mK |
| | | | | | 1.22 µA (0.006% of full scale) | | | | 0.8 mK | | 1.5 mK | | 7.6 mK | | 15.3 mK | | 22.9 mK | | 19.1 mK |
|
| Output accuracy | | Current | | Temperature equivalence | | Range 1 | | Range 2 | | Range 3 | | Range 4 | | Range 5 | | Range 6 |
| | | | ±2 µA (±0.01% of full scale) | | | | 2.5 mK | | 12.5 mK | | 25 mK | | 41 mK | | 59 mK | | 125 mK |
| | | | ±2 µA (±0.01% of full scale) | | | | Not used | | 12.5 mK | | 25 mK | | 41 mK | | 59 mK | | 125 mK |
| | | | ±5 µA (±0.025% of full scale) | | | | 3.1 mK | | 6.2 mK | | 31.2 mK | | 62.5 mK | | 93.7 mK | | 78.1 mK |
|
| Output temperature coefficient c
| | Current (%/°C ambient) | | | | Temperature equivalence | | Range 1 | | Range 2 | | Range 3 | | Range 4 | | Range 5 | | Range 6 |
| | | | ±0.0055% of output current | | per °C | | | | 1 mK/°C | | 6 mK/°C | | 12 mK/°C | | 18 mK/°C | | 26 mK/°C | | 55 mK/°C |
| | | | ±0.0055% of output current | | per °C | | | | Not used | | 6 mK/°C | | 12 mK/°C | | 18 mK/°C | | 26 mK/°C | | 55 mK/°C |
| | | | ±2 µA/°C (±0.01%/°C) | | | | | | ±1 mK/°C | | ±2 mK/°C | | ±10 mK/°C | | ±20 mK/°C | | ±30 mK/°C | | ±25 mK/°C |
|
| 0 mA to 20 mA output (0 V to 10 V with
| | 500 Ω, 0.02% load resistor) | | Output resolution | | Voltage | | Temperature equivalence | | Range 1 | | Range 2 | | Range 3 | | Range 4 | | Range 5 | | Range 6 |
| | | | | | | | 0.6 mV | | | | 1.2 mK | | 6.1 mK | | 12.2 mK | | 19.8 mK | | 29 mK | | 61 mK |
| | | | | | | | 0.6 mV | | | | Not used | | 6.1 mK | | 12.2 mK | | 19.8 mK | | 29 mK | | 61 mK |
| | | | | | | | 0.61 mV | | | | 0.6 mK | | 1.2 mK | | 6.1 mK | | 12.2 mK | | 18.3 mK | | 15.2 mK |
|
| Output accuracy
| | Voltage | | | | Temperature equivalence | | Range 1 | | Range 2 | | Range 3 | | Range 4 | | Range 5 | | Range 6 |
| | | | 3 mV (0.03% of full scale) | | | | | | 6 mK | | 30 mK | | 60 mK | | 98 mK | | 143 mK | | 300 mK |
| | | | 3 mV (0.03% of full scale) | | | | | | Not used | | 30 mK | | 60 mK | | 98 mK | | 143 mK | | 300 mK |
| | | | ±4.5 mV (±0.025% of full scale | | ±0.02% resistor accuracy) | | | | 4.5 mK | | 9.0 mK | | 45.0 mK | | 90.0 mK | | 135 mK | | 112.5 mK |
|
| Output temperature coefficient
| | Voltage (% output/°C ambient) | | | | Temperature equivalence | | Range 1 | | Range 2 | | Range 3 | | Range 4 | | Range 5 | | Range 6 |
| | | | ±0.008%/°C | | | | | | 2 mK/°C | | 8 mK/°C | | 16 mK/°C | | 26 mK/°C | | 38 mK/°C | | 80 mK/°C |
| | | | ±0.008%/°C | | | | | | Not used | | 8 mK/°C | | 16 mK/°C | | 26 mK/°C | | 38 mK/°C | | 80 mK/°C |
| | | | ±1.25 mV/°C (±0.01%/°C | | ±0.0025%/°C of load resistor) | | | | ±1.2 mK/°C | | ±2.5 mK/°C | | ±12 mK/°C | | ±25 mK/°C | | ±36 mK/°C | | ±30 mK/°C |
|
General |
| Ambient temperature range | 15 °C to 35 °C | 15 °C to 35 °C | 15 °C to 35 °C |
| Power requirements | ±5 (±0.25) VDC, (500 mA), 2.5 W | ±5 (±0.25) VDC, (500 mA), 2.5 W | 234: ±5 (±0.25) VDC, (500 mA), 2.5 W |
| Enclosure type | see diagrams | see diagrams | see diagrams |
| Mounting | VME end panel and back plane: transmitters do not use electrical bus format, only its physical shape and power | VME end panel and back plane: transmitters do not use electrical bus format, only its physical shape and power | VME end panel and back plane: transmitters do not use electrical bus format, only its physical shape and power |
| Size | 100 mm H × 160 mm D × 30.5 mm W | 100 mm H × 160 mm D × 30.5 mm W | 234: 100 mm H × 160 mm D × 30.5 mm W |
| Weight | 0.25 kg (0.5 lb) | 0.25 kg (0.5lb) | 0.25 kg (0.5 lb) |