Click here for sample calculations of typical sensor performance
 
Example Lake Shore sensor Temp (K) Nominal
resistance/
voltage
Typical
sensor sensitivity4
Measurement
resolution:
temperature
equivalents
Electronic accuracy: temperature equivalents Temperature accuracy including electronic accuracy, CalCurve™, and calibrated sensor Electronic control stability5:
temperature equivalents
Cernox™
(1 mV)
CX-1010-SD with 0.1L calibration 0.1 21389 Ω -558110 Ω/K 5.4 µK ±69 µK ±4.1 mK ±10.8 µK
0.3 2322.4 Ω -10785 Ω/K 28 µK ±272 µK ±4.3 mK ±55.6 µK
0.5 1248.2 Ω -2665.2 Ω/K 113 µK ±938 µK ±4.9 mK ±225 µK
4.2 277.32 Ω -32.209 Ω/K 931 µK ±6.5 mK ±11 mK ±190 µK
300 30.392 Ω -0.0654 Ω/K 153 mK ±1.7 K ±1.8 K ±306 mK
Cernox™
(10 mV)
CX-1050-SD-HT4 with 1.4M calibration 1.4 26566 Ω -48449 kΩ/K 6.2 µK ±261 µK ±4.3 mK ±12.4 µK
4.2 3507.2 Ω -1120.8 kΩ/K 89 µK ±2.1 mK ±6.1 mK ±178 µK
77 205.67 Ω -2.4116 Ω/K 1.2 mK ±38 mK ±50 mK ±2.5 mK
420 45.03 Ω -0.0829 Ω/K 12 mK ±338 mK ±412 mK ±24.1 mK
Germanium
(1 mV)
GR-50-AA with 0.05A calibration 0.1 2317 Ω -71858 Ω/K 4.2 µK ±21 µK ±4.0 mK ±8.3 µK
0.3 164 Ω -964 Ω/K 31.1 µK ±130 µK ±4.1 mK ±62.2 µK
0.5 73.8 Ω -202.9 Ω/K 49.3 µK ±244 µK ±4.2 mK ±98.6 µK
1.4 24.7 Ω -13.15 Ω/K 228 µK ±1.2 mK ±5.2 mK ±456 µK
4.2 13.7 Ω -1.036 Ω/K 2.9 mK ±11 mK ±15 mK ±5.8 mK
Germanium
(10 mV)
GR-300-AA with 0.3D calibration 0.3 35180 Ω -512200 Ω/K 2 µK ±47 µK ±4.0 mK ±3.9 µK
1.4 448.6 Ω -581.3 Ω/K 17 µK ±481 µK ±4.5 mK ±34.4 µK
4.2 94.46 Ω -26.56 Ω/K 38 µK ±1.8 mK ±5.8 mK ±75.3 µK
100 2.72 Ω -0.024 Ω/K 4.2 mK ±151 mK ±171 mK ±8.3 mK
Germanium
(10 mV)
GR-1400-AA with 1.4D calibration 1.4 35890 Ω -94790 Ω/K 11 µK ±257 µK ±4.3 mK ±21.1 µK
4.2 1689 Ω -861.9 Ω/K 35 µK ±900 µK ±4.9 mK ±69.6 µK
77 3.55 Ω -0.05 Ω/K 2 mK ±83 mK ±94 mK ±4 mK
100 2.8 Ω -0.021 Ω/K 4.8 mK ±175 mK ±195 mK ±9.5 mK
Rox™
(1 mV)
RX-102B-CB with 0.02C calibration 0.1 3549 Ω -12578 Ω/K 79.5 µK ±908 µK ±4.9 mK ±159 µK
0.5 2188 Ω -1056 Ω/K 284 µK ±2.7 mK ±6.7 mK ±568 µK
1.4 1779 Ω -198 Ω/K 1.5 mK ±13.7 mK ±18 mK ±3.0 mK
4.2 1546 Ω -40.0 Ω/K 7.5 mK ±65.4 mK ±69 mK ±15.0 mK
40 1199 Ω -3.41 Ω/K 88 mK ±727 mK ±803 mK ±176 mK
Platinum RTD
500 Ω Full Scale
PT-103
with 14J calibration
30 3.66 Ω 0.191 Ω/K 0.5 mK ±22 mK ±31 mK ±1.0 mK
77 20.38 Ω 0.423 Ω/K 0.7 mK ±34 mK ±44 mK ±1.4 mK
300 110.35 Ω 0.387 Ω/K 7.8 mK ±140 mK ±164 mK ±15.5 mK
500 185.668 Ω 0.378 Ω/K 7.9 mK ±223 mK ±274 mK ±15.9 mK
Silicon Diode     DT-670-C0-13 with 1.4H calibration     1.4 1.664 V -12.49 mV/K 0.8 mK ±13 mK ±20 mK ±1.6 mK
77 1.028 V -1.73 mV/K 5.8 mK ±76 mK ±113 mK ±11.6 mK
300 0.5596 V -2.3 mV/K 4.3 mK ±47 mK ±82 mK ±8.7 mK
500 0.0907 V -2.12 mV/K 4.7 mK ±40 mK ±94 mK ±9.4 mK
Silicon Diode DT-470-SD-13 with 1.4H calibration 1.4 1.6981 V -13.1 mV/K 0.8 mK ±13 mK ±20 mK ±1.5 mK
77 1.0203 V -1.92 mV/K 5.2 mK ±68 mK ±105 mK ±10.4 mK
300 0.5189 V -2.4 mV/K 4.2 mK ±44 mK ±79 mK ±8.3 mK
475 0.0906 V -2.22 mV/K 4.5 mK ±38 mK ±87 mK ±9.0 mK
GaAlAs Diode TG-120-SD with 1.4H calibration 1.4 5.3909 V -97.5 mV/K 0.21 mK ±6 mK ±13 mK ±410 µK
77 1.4222 V -1.24 mV/K 16 mK ±179 mK ±216 mK ±32.3 mK
300 0.8978 V -2.85 mV/K 7 mK ±60 mK ±95 mK ±14.0 mK
475 0.3778 V -3.15 mV/K 6.3 mK ±37 mK ±86 mK ±12.7 mK
Capacitance CS-501 4.2 6.0 nF 27 pF/K 7.4 mK Not applicable
Calibration not available from
Lake Shore
±14.8 mK
77 9.1 nF 52 pF/K 3.9 mK ±7.7 mK
200 19.2 nF 174 pF/K 12 mK ±23 mK
Thermocouple
50 mV
3060-F
Type K 75 -5862.9 µV 15.6 µV/K 26 mK ±252 mK5
Calibration not available from
Lake Shore
   
±51.3 mK
300 1075.3 µV 40.6 µV/K 9.9 mK ±38 mK5 ±19.7 mK
600 13325 µV 41.7 µV/K 9.6 mK ±184 mK5 ±19.2 mK
1500 49813 µV 36.1 µV/K 11 mK ±718 mK5 ±22.2 mK
2 Typical sensor sensitivities were taken from representative calibrations for the sensor listed
3 Control stability of the electronics only, in an ideal thermal system
4 Non-HT version maximum temperature: 325 K
5 Accuracy specification does not include errors from room temperature compensation