Thermometry |
| Number of inputs |
2 |
Input configuration |
Inputs can be configured from the front panel to accept any of the supported input types. Thermocouple inputs require an optional input card that can be installed in the field. Once installed the thermocouple input can be selected from the front panel like any other input type. |
Isolation |
Sensor inputs optically isolated from other circuits but not each other |
A/D resolution |
24-bit |
Input accuracy |
Sensor dependent, refer to Input Specifications table |
Measurement resolution |
Sensor dependent, refer to Input Specifications table |
Maximum update rate |
10 rdg/s on each input, 5 rdg/s when configured as 100 kΩ NTC RTD with reversal on |
Autorange |
Automatically selects appropriate NTC RTD or PTC RTD range |
User curves |
Room for 39 200-point CalCurves™ or user curves |
SoftCal™ |
Improves accuracy of DT-470 diode to ±0.25 K from 30 K to 375 K; improves accuracy of platinum RTDs to ±0.25 K from 70 K to 325 K; stored as user curves |
Math |
Maximum and minimum |
Filter |
Averages 2 to 64 input readings |
|
|
Sensor Input Configuration |
|
Diode/RTD |
Thermocouple |
| Measurement type |
4-lead differential |
2-lead differential, room temperature compensated |
| Excitation |
Constant current with current reversal for RTDs |
NA |
Supported sensors |
Diodes: Silicon, GaAlAs
RTDs: 100 Ω Platinum, 1000 Ω Platinum, Germanium, Carbon-Glass, Cernox™, and Rox™ |
Most thermocouple types |
Standard curves |
DT-470, DT-670, DT-500D, DT-500-E1, PT-100, PT-1000, RX-102A, RX-202A |
Type E, Type K, Type T, AuFe 0.07% vs. Cr,
AuFe 0.03% vs Cr, |
Input connector |
6-pin DIN |
Screw terminals in a ceramic isothermal block |
|
Control |
|
| Control outputs |
2 |
|
|
Heater Outputs |
|
Control type
|
Closed loop digital PID with manual heater output or open loop; warm up mode (output 2 only) |
Update rate
|
10/s |
Tuning
|
Autotune (one loop at a time), PID, PID zones |
Control stability
|
Sensor dependent – see Input Specifications table |
PID control parameters
|
| Proportional (gain) |
0 to 1000 with 0.1 setting resolution |
| Integral (reset) |
1 to 1000 (1000/s) with 0.1 setting resolution |
| Derivative (rate) |
1 to 200% with 1% resolution |
Manual output
|
0 to 100% with 0.01% setting resolution |
Zone control
|
10 temperature zones with P, I, D, manual heater out, heater range, control channel, ramp rate |
| Setpoint ramping |
0.1 K/min to 100 K/min |
|
|
|
|
Output 1 |
| Type |
Variable DC current source |
| Control modes |
Closed loop digital PID with manual output or open loop |
| D/A resolution |
16-bit |
|
25 Ω setting |
50 Ω setting |
| Max power |
75 W* |
50 W |
50 W |
| Max current |
1.73 A |
1.41 A |
1 A |
| Voltage compliance (min) |
43.3 V |
35.4 V |
50 V |
| Heater load for max power |
25 Ω |
25 Ω |
50 Ω |
| Heater load range |
10 Ω to 100 Ω |
| Ranges |
3 (decade steps in power) |
| Heater noise |
0.12 µA RMS (dominated by line frequency and its harmonics) |
| Heater connector |
Dual banana |
| Grounding |
Output referenced to chassis ground |
| Safety limits |
Curve temperature, power up heater off, short circuit protection |
| *75 W only available when output 2 is in voltage mode |
|
|
Output 2 |
| Type |
Variable DC current source or voltage source |
|
Current mode |
Voltage mode |
| Control modes |
Closed loop digital PID with manual output, zone, open loop |
Closed loop digital PID with manual output, zone, open loop, warm up, monitor out |
| D/A resolution |
15-bit |
16-bit (bipolar)/15-bit (unipolar) |
|
25 Ω setting |
50 Ω setting |
NA |
| Max power |
25 W |
25 W |
1 W |
| Max current |
1 A |
0.71 A |
100 mA |
| Voltage compliance (min) |
25 V |
35.4 V |
±10 V |
| Heater load for max power |
25 Ω |
50 Ω |
100 Ω |
| Heater load range |
10 Ω to 100 Ω |
100 Ω min (short circuit protected) |
| Ranges |
3 (decade steps in power) |
NA |
| Heater noise |
0.12 µA RMS |
0.3 mV RMS |
| Heater connector |
Dual banana |
Detachable terminal block |
| Grounding |
Output referenced to chassis ground |
| Safety limits |
Curve temperature, power up heater off, short circuit protection |
|
|
Warm up heater mode settings (Output 2 voltage only)
|
| Warm up percentage |
0 to 100% with 1% resolution |
| Warm up mode |
Continuous control or auto-off |
Monitor output settings (Output 2 voltage only)
|
| Scale |
User selected |
| Data source |
Temperature or sensor units |
| Settings |
Input, source, top of scale, bottom of scale, or manual |
| Update rate |
10/s |
| Range |
±10 V |
| Resolution |
16-bit, 0.3 mV |
| Accuracy |
±2.5 mV |
| Noise |
0.3 mV RMS |
| Minimum load resistance |
100 Ω (short circuit protected) |
| Connector |
Detachable terminal block |
Front Panel |
|
Display
|
2- line by 20-character, 9 mm character height, vacuum fluorescent display |
Number of reading displays
|
1 to 4 |
Display units
|
K, °C, V, mV, Ω |
Reading source
|
Temperature, sensor units, max and min |
Display update rate
|
2 rdg/s |
Temp display resolution
|
0.001° from 0° to 99.999°, 0.01° from 100° to 999.99°, 0.1° above 1000° |
Sensor units display resolution
|
Sensor dependent, to 5 digits |
Other displays
|
Sensor name, setpoint, heater range, heater output, and PID |
Setpoint setting resolution
|
Same as display resolution (actual resolution in sensor dependent) |
Heater output display
|
Numeric display in percent of full scale for power or current |
Heater output resolution
|
1% |
Display annunciators
|
Control input, alarm, tuning |
LED annunciators
|
Remote, alarm, control outputs |
Keypad
|
25-key silicone elastomer keypad |
| Front panel features |
Front panel curve entry, display brightness control, and keypad lock-out |
|
|
|
Interface |
|
IEEE-488.2
|
| Capabilities |
SH1, AH1, T5, L4, SR1, RL1, PP0, DC1, DT0, C0, E1 |
| Reading rate |
To 10 rdg/s on each input |
Software support
|
LabVIEW™ driver (consult Lake Shore for availability) |
USB
|
| Function |
Emulates a standard RS-232 serial port |
| Baud rate |
57,600 |
| Connector |
B-type USB connector |
| Reading rate |
To 10 rdgs/s on each input |
| Software support |
LabVIEW™ driver (consult Lake Shore for availability) |
Special interface features
|
Model 331/332 command emulation mode |
Alarms
|
| Number |
2, high and low for each input |
| Data source |
Temperature or sensor units |
| Settings |
Source, high setpoint, low setpoint, deadband, latching or non-latching, audible on/off, and visible on/off |
Actuators |
Display annunciator, beeper, and relays |
Relays
|
| Number |
2 |
| Contacts |
Normally open (NO), normally closed (NC), and common (C) |
| Contact rating |
30 VDC at 3 A |
| Operation |
Activate relays on high, low, or both alarms for any input, or manual mode |
| Connector |
Detachable terminal block |
|
|
|
General |
|
Ambient temperature
|
15 °C to 35 °C at rated accuracy, 5 °C to 40 °C at reduced accuracy |
Power requirement
|
100, 120, 220, 240 VAC, ±10%, 50 or 60 Hz, 210 VA |
Size
|
217 mm W × 90 mm H × 317 mm D (8.5 in × 3.5 in × 14.5 in), half rack |
Weight
|
5.1 kg (11.3 lb) |
| Approval |
CE mark |
|
|
|
|
|
|